Homemade valved holding chambers for children with airway hyperresponsiveness: A randomized crossover trial
Kesanee Chaicoming, อรุณวรรณ พฤทธิพันธุ์*, Anusorn Adirekkittikun, Malinee Nugboon
Division of Pediatric Pulmonology, Department of Pediatrics, Faculty of Medicine, Ramathibodi Hospital, Mahidol University, Bangkok 10400, Thailand; Email: [email protected]
บทคัดย่อ

Background: During the COVID-19 pandemic, a metered-dose inhaler (MDI) with a valved holding chamber (VHC) is a preferred route of bronchodilator delivery. We have developed a new homemade VHC, made of a paper coffee cup, and a drinking water bottle. This study was conducted to compare the bronchodilator response in children with airway hyperresponsiveness after the use of our homemade VHC and that of a standard commercial one.

Methods: In a randomized, two-period, two-sequence crossover trial, we recruited 20 children, aged 6-15 years, who had a greater than 12% increase in FEV1 after inhaled salbutamol. They were randomized into Group A and B. Group A used our VHC on the first day and Aerochamber® on the second day. Group B used the same VHCs but in alternate sequence. Spirometries were performed before and after 400 µg of salbutamol, MDI was administered via those VHCs.

Results: Baseline demographic data and spirometric values did not have statistically significant differences between group A and B and between the first and second day (p > .05). After giving salbutamol MDI, both VHCs produced significant increases in FVC, FEV1 , and FEF25-75% (p < .005). The improvement in FEV1 did not significantly differ between our homemade VHC and Aerochamber® (p > .05).

Conclusion: Our homemade VHC is effective for an MDI bronchodilator delivery. Since it is very cheap and easy to make, it may be used as a disposable device to minimize airborne transmission especially when commercial VHC is not available.

 
ที่มา
Pediatric Pulmonology ปี 2564, January ปีที่: 56 ฉบับที่ 1 หน้า
คำสำคัญ
49-56